Search found 2 matches

by gfacheris
Fri Apr 04, 2025 1:01 pm
Forum: Particle number size distribution – mobility diameter 10 to 800 nm
Topic: Pulse Height Error on Pulse Height CPC 3750
Replies: 1
Views: 29457

Re: Pulse Height Error on Pulse Height CPC 3750

UPDATE

I managed to solve the pulse height issue by completely dismounting the optics and cleaning the optical chamber, the laser and the optics.
In my case I did not had to realign the laser by means of an oscilloscope but it could be the case.
by gfacheris
Tue Jan 14, 2025 12:53 pm
Forum: Particle number size distribution – mobility diameter 10 to 800 nm
Topic: Pulse Height Error on Pulse Height CPC 3750
Replies: 1
Views: 29457

Pulse Height Error on Pulse Height CPC 3750

Dear All, I have the following issue on the DMPS. Since few weeks the Total CPC (TSI 3750) is returning an error/warning related to pulse height value. The value itself is varying during the day from very low values (30% to 80%) and from time to time the error acknowledges itself (see attached image...